Caution on the use of NBS 30 biotite for hydrogen-isotope measurements with on-line high-temperature conversion systems
Abstract
RATIONALE The supply of NBS 30 biotite is nearly exhausted. During measurements of NBS 30 and potential replacements, reproducible δ 2 H VSMOW-SLAP values could not be obtained by three laboratories using high-temperature conversion (HTC) systems. The cause of this issue has been investigated using the silver-tube technique for hydrogen-isotope measurements of water. METHODS The δ 2 H VSMOW-SLAP values of NBS 30 biotite, other biotites, muscovites, and kaolinite with different particle sizes, along with IAEA-CH-7 polyethylene, and reference waters and NBS 22 oil that were sealed in silver-tube segments, were measured. The effect of absorbed water on mineral surfaces was investigated with waters both enriched and depleted in 2 H. The quantitative conversion of hydrogen from biotite into gaseous hydrogen as a function of mass and particle size was also investigated. RESULTS The δ 2 H VSMOW-SLAP values of NBS 30 obtained by three laboratories were as much as 21 ‰ too high compared with the accepted value of −65.7 ‰, determined by conventional off-line measurements. The experiments showed a strong correlation between grain size and the δ 2 H VSMOW-SLAP value of NBS 30 biotite, but not of biotites with lower iron content. The δ 2 H VSMOW-SLAP values of NBS 30 as a function of particle size show a clear trend toward −65.7 ‰ with finer grain size. CONCLUSIONS Determination of the δ 2 H VSMOW-SLAP values of hydrous minerals and of NBS 30 biotite by on-line HTC systems coupled to isotope-ratio mass spectrometers may be unreliable because hydrogen in this biotite may not be converted quantitatively into molecular hydrogen. Extreme caution in the use and interpretation of δ 2 H VSMOW-SLAP on-line measurements of hydrous minerals is recommended.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Haiping Qi, Tyler B. Coplen, Gerard Olack, Torsten W. Vennemann. 2014-08-05. Caution on the use of NBS 30 biotite for hydrogen-isotope measurements with on-line high-temperature conversion systems. https://doi.org/10.1002/rcm.6983
Cite the original work for its findings. Save a collection to share your selection of sources.