Quality-Assurance Data for Routine Water Analyses by the U.S. Geological Survey Laboratory in Troy, New York--July 1999 through June 2001

Abstract

The laboratory for analysis of low-ionic-strength water at the U.S. Geological Survey (USGS) Water Science Center in Troy, N.Y., analyzes samples collected by USGS projects throughout the Northeast. The laboratory's quality-assurance program is based on internal and interlaboratory quality-assurance samples and quality-control procedures that were developed to ensure proper sample collection, processing, and analysis. The quality-assurance and quality-control data were stored in the laboratory's LabMaster data-management system, which provides efficient review, compilation, and plotting of data. This report presents and discusses results of quality-assurance and quality-control samples analyzed from July 1999 through June 2001. Results for the quality-control samples for 18 analytical procedures were evaluated for bias and precision. Control charts indicate that data for eight of the analytical procedures were occasionally biased for either high-concentration or low-concentration samples but were within control limits; these procedures were: acid-neutralizing capacity, total monomeric aluminum, total aluminum, calcium, chloride and nitrate (ion chromatography and colormetric method) and sulfate. The total aluminum and dissolved organic carbon procedures were biased throughout the analysis period for the high-concentration sample, but were within control limits. The calcium and specific conductance procedures were biased throughout the analysis period for the low-concentration sample, but were within control limits. The magnesium procedure was biased for the high-concentration and low concentration samples, but was within control limits. Results from the filter-blank and analytical-blank analyses indicate that the procedures for 14 of 15 analytes were within control limits, although the concentrations for blanks were occasionally outside the control limits. The data-quality objective was not met for dissolved organic carbon. Sampling and analysis precision are evaluated herein in terms of the coefficient of variation obtained for triplicate samples in the procedures for 17 of the 18 analytes. At least 90 percent of the samples met data-quality objectives for all analytes except ammonium (81 percent of samples met objectives), chloride (75 percent of samples met objectives), and sodium (86 percent of samples met objectives). Results of the USGS interlaboratory Standard Reference Sample (SRS) Project indicated good data quality over the time period, with most ratings for each sample in the good to excellent range. The P-sample (low-ionic-strength constituents) analysis had one satisfactory rating for the specific conductance procedure in one study. The T-sample (trace constituents) analysis had one satisfactory rating for the aluminum procedure in one study and one unsatisfactory rating for the sodium procedure in another. The remainder of the samples had good or excellent ratings for each study. Results of Environment Canada's National Water Research Institute (NWRI) program indicated that at least 89 percent of the samples met data-quality objectives for 10 of the 14 analytes; the exceptions were ammonium, total aluminum, dissolved organic carbon, and sodium. Results indicate a positive bias for the ammonium procedure in all studies. Data-quality objectives were not met in 50 percent of samples analyzed for total aluminum, 38 percent of samples analyzed for dissolved organic carbon, and 27 percent of samples analyzed for sodium. Results from blind reference-sample analyses indicated that data-quality objectives were met by at least 91 percent of the samples analyzed for calcium, chloride, fluoride, magnesium, pH, potassium, and sulfate. Data-quality objectives were met by 75 percent of the samples analyzed for sodium and 58 percent of the samples analyzed for specific conductance.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Tricia A. Lincoln, Debra A. Horan-Ross, Michael R. McHale, Gregory B. Lawrence. 2006. Quality-Assurance Data for Routine Water Analyses by the U.S. Geological Survey Laboratory in Troy, New York--July 1999 through June 2001. https://doi.org/10.3133/ofr20061246

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related USGS reports

ECCOE Landsat quarterly calibration and validation report—Quarter 2, 2026

Executive Summary The U.S. Geological Survey Earth Resources Observation and Science Calibration and Validation (Cal/Val) Center of Excellence (ECCOE) focuses on improving the accuracy, precision, calibration, and product quality of remote-sensing data, leveraging years of multiscale optical system geometric and radiometric calibration and characterization experience. The ECCOE Landsat Cal/Val Team continually monitors the geometric and radiometric performance of active Landsat missions and makes calibration adjustments, as needed, to maintain data quality at the highest level. This report provides observed geometric and radiometric analysis results for Landsats 8 and 9 for quarter 2 (April–June), 2026. All data used to compile the Cal/Val analysis results presented in this report are freely available from the U.S. Geological Survey EarthExplorer website: https://earthexplorer.usgs.gov.

Open-File Report↗

ECCOE Landsat quarterly calibration and validation report—Quarter 1, 2026

Executive Summary The U.S. Geological Survey Earth Resources Observation and Science Calibration and Validation (Cal/Val) Center of Excellence (ECCOE) focuses on improving the accuracy, precision, calibration, and product quality of remote-sensing data, leveraging years of multiscale optical system geometric and radiometric calibration and characterization experience. The ECCOE Landsat Cal/Val Team continually monitors the geometric and radiometric performance of active Landsat missions and makes calibration adjustments, as needed, to maintain data quality at the highest level. This report provides observed geometric and radiometric analysis results for Landsats 8 and 9 for quarter 1 (January–March) of 2026. All data used to compile the Cal/Val analysis results presented in this report are freely available from the U.S. Geological Survey EarthExplorer website: https://earthexplorer.usgs.gov . One specific activity that the ECCOE Landsat Cal/Val Team closely monitored was a Landsat 8 Thermal Infrared Sensor encoder alerts anomaly. On March 18, 2026, the Thermal Infrared Sensor instrument’s mechanism control electronics board reset, driving several parameters to default power-on values. Subsequent telemetry and analysis determined that no persistent hardware misconfiguration was present and no radiometric or geometric degradation was observed.

Open-File Report↗

Floods of June 20–July 6, 2024, in the Big Sioux River, Rock River, Little Sioux River, Ocheyedan River, and Floyd River Basins, northwestern Iowa

Major flooding occurred on June 20–July 6, 2024, in northwestern Iowa affecting the Big Sioux, Rock, Little Sioux, Ocheyedan, and Floyd River Basins. Heavy rain fell in northwestern Iowa, southwestern Minnesota, and southeastern South Dakota on June 20–22, 2024. Parts of northwestern Iowa recorded 2–6 inches of rainfall and localized amounts exceeding 12 inches. A maximum peak-of-record streamflow of 175,000 cubic feet per second at the U.S. Geological Survey streamgage Big Sioux River at Akron, Iowa (06485500), was recorded on June 22, 2024, and had an annual exceedance probability range of 0.2–0.49 percent. High-water marks were measured at four locations along the Big Sioux River between U.S. Interstate 29 at Sioux City, Iowa, upstream to Iowa Highway 10 north of Hawarden, Iowa, a distance of 75.7 river miles. A maximum peak-of-record streamflow of 157,000 cubic feet per second at the U.S. Geological Survey streamgage Rock River near Rock Valley, Iowa (06483500), was recorded on June 22, 2024, and had an annual exceedance probability of less than 0.2 percent. High-water marks were measured at eight locations along the Rock River between County Road B30 east of Hudson, South Dakota, upstream to Iowa Highway 9 at Rock Rapids, Iowa, a distance of 39.3 river miles. A maximum peak-of-record streamflow of 63,000 cubic feet per second at the U.S. Geological Survey streamgage Little Sioux River at Correctionville, Iowa (06606600), was recorded on June 24, 2024, and had an annual exceedance probability range of 0.2–0.49 percent. High-water marks were measured at 11 locations along the Little Sioux River between Iowa Highway 31 west of Correctionville, Iowa, upstream to U.S. Highway 18 north of Spencer, Iowa, a distance of 134.8 river miles. A maximum streamflow of 24,500 cubic feet per second at the U.S. Geological Survey streamgage Ocheyedan River near Spencer, Iowa (06605000), was recorded on June 22, 2024, and had an annual exceedance probability range of 0.2–0.49 percent. High-water marks were measured at three locations along the Ocheyedan River between County Road M38 west of Spencer, Iowa, upstream to U.S. Highway 18 west of Everly, Iowa, a distance of 12.8 river miles. A maximum streamflow of 41,000 cubic feet per second at the U.S. Geological Survey streamgage Floyd River at Alton, Iowa (06600100), was recorded on June 22, 2024, and had an annual exceedance probability range of 1–1.99 percent. High-water marks were measured at six locations along the Floyd River between Iowa Highway 3 at Le Mars, Iowa, upstream to Iowa Highway 10 at Alton, Iowa, a distance of 27.5 river miles. The high-water marks were used to develop flood profiles for the Big Sioux, Rock, Little Sioux, Ocheyedan, and Floyd Rivers.

Iowa, Minnesota, South Dakota↗