Radioisotope dilution analyses of geological samples using 236U and 229Th

Abstract

The use of 236U and 229Th in alpha spectrometric measurements has some advantages over the use of other tracers and measurement techniques in isotope dilution analyses of most geological samples. The advantages are: (1) these isotopes do not occur in terrestrial rocks, (2) they have negligible decay losses because of their long half lives, (3) they cause minimal recoil contamination to surface-barrier detectors, (4) they allow for simultaneous determination of the concentration and isotopic composition of uranium and thorium in a variety of sample types, and (5) they allow for simple and constant corrections for spectral inferences, 0.5% of the 238U activity is subtracted for the contribution of 235U in the 236U peak and 1% of the 229Th activity is subtracted from the 230Th activity. Disadvantages in using 236U and 229Th are: (1) individual separates of uranium and thorium must be prepared as very thin sources for alpha spectrometry, (2) good resolution in the spectrometer system is required for thorium isotopic measurements where measurement times may extend to 300 h, and (3) separate calibrations of the 236U and 229Th spike solution with both uranium and thorium standards are required. The use of these tracers in applications of uranium-series disequilibrium studies has simplified the measurements required for the determination of the isotopic composition of uranium and thorium because of the minimal corrections needed for alpha spectral interferences. ?? 1984.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J.N. Rosholt. 1984. Radioisotope dilution analyses of geological samples using 236U and 229Th. https://doi.org/10.1016/0167-5087(84)90711-7

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related USGS reports

Inter-pulse high-resolution gamma-ray spectra using a 14 MeV pulsed neutron generator

A neutron generator pulsed at 100 s-1 was suspended in an artificial borehole containing a 7.7 metric ton mixture of sand, aragonite, magnetite, sulfur, and salt. Two Ge(HP) gamma-ray detectors were used: one in a borehole sonde, and one at the outside wall of the sample tank opposite the neutron generator target. Gamma-ray spectra were collected by the outside detector during each of 10 discrete time windows during the 10 ms period following the onset of gamma-ray build-up after each neutron burst. The sample was measured first when dry and then when saturated with water. In the dry sample, gamma rays due to inelastic neutron scattering, neutron capture, and decay were counted during the first (150 ??s) time window. Subsequently only capture and decay gamma rays were observed. In the wet sample, only neutron capture and decay gamma rays were observed. Neutron capture gamma rays dominated the spectrum during the period from 150 to 400 ??s after the neutron burst in both samples, but decreased with time much more rapidly in the wet sample. A signal-to-noise-ratio (S/N) analysis indicates that optimum conditions for neutron capture analysis occurred in the 350-800 ??s window. A poor S/N in the first 100-150 ??s is due to a large background continuum during the first time interval. Time gating can be used to enhance gamma-ray spectra, depending on the nuclides in the target material and the reactions needed to produce them, and should improve the sensitivity of in situ well logging. ?? 1984.

Nuclear Instruments and Methods In Physics Researc