Reconnaissance for mercury over geothermal areas of the Imperial Valley, California
No abstract available.
Geology topics
Publications and source records attributed to William W. Vaughn.
No abstract available.
The detector utilizes a large-volume atomic-absorption technique for quantitative determinations of mercury vapor thermally released from crushed rock. A quartz-enclosed noble-metal amalgamative stage, which is temperature controlled and is actuated by a radio-frequency induction heater, selectively traps the mercury and eliminates low-level contamination. As little as 1 part per billion of mercury can be detected in a 1-gram sample in a 1-minute analytical period.
A low-cost instrument has been developed for quantitative determinations of beryllium in the field by activation analysis. The instrument makes use of the gamma-neutron reaction between gammas emitted by an artificially radioactive source (Sb124) and beryllium as it occurs in nature. The instrument and power source are mounted in a panel-type vehicle. Samples are prepared by hand-crushing the rock to approximately ?-inch mesh size and smaller. Sample volumes are kept constant by means of a standard measuring cup. Instrument calibration, made by using standards of known BeO content, indicates the analyses are reproducible and accurate to within ? 0.25 percent BeO in the range from 1 to 20 percent BeO with a sample counting time of 5 minutes. Sensitivity of the instrument maybe increased somewhat by increasing the source size, the sample size, or by enlarging the cross-sectional area of the neutron-sensitive phosphor normal to the neutron flux.