Chapter A5. Section 6.4.B. Low-Level Mercury

Collecting and processing water samples for analysis of mercury at a low (subnanogram per liter) level requires use of ultratrace-level techniques for equipment cleaning, sample collection, and sample processing. Established techniques and associated quality-assurance (QA) procedures for the collection and processing of water samples for trace-element analysis at the part-per-billion level (NFM 3-5) are not adequate for low-level mercury samples. Modifications to the part-per-billion procedures are necessary to minimize contamination of samples at a typical ambient mercury concentration, which commonly is at the subnanogram-per-liter level.

Techniques of Water-Resources Investigations

Chapter A5. Section 6.1.F. Wastewater, Pharmaceutical, and Antibiotic Compounds

The USGS differentiates between samples collected for analysis of wastewater compounds and those collected for analysis of pharmaceutical and antibiotic compounds, based on the analytical schedule for the laboratory method. Currently, only the wastewater laboratory method for field-filtered samples (SH1433) is an approved, routine (production) method. (The unfiltered wastewater method LC 8033 also is available but requires a proposal for custom analysis.) At this time, analysis of samples for pharmaceutical and antibiotic compounds is confined to research studies and is available only on a custom basis.

Techniques of Water-Resources Investigations