On-orbit performance of the Landsat-7 ETM+ radiometric calibrators

The Landsat-7 Enhanced Thematic Mapper Plus (ETM+) incorporates two new devices to improve its absolute radiometric calibration: a Full Aperture Solar Calibrator (FASC) and a Partial Aperture Solar Calibrator (PASC). The FASC is a diffuser panel, typically deployed once per month. Initial FASC absolute calibration results were within 5% of the pre-launch calibrations. Over time, the responses of the ETM+ to the FASC have varied with the location viewed on the panel, suggesting a localized degradation or contamination of the panel. On the best part of the panel, the trends in response range from m 1.4% y m 1 (band 4) to +0.6% y m 1 (band 7), with band 5 showing the least change at m 0.4% y m 1 . Changes in the panel reflectance due to UV exposure are believed to be the origin of these trends. The PASC is a set of auxiliary optics that allows the ETM+ to image the Sun through reduced apertures. PASC data have normally been acquired on a daily basis. Unlike the FASC, the PASC has exhibited significant anomalies. During the first six months of operation, responses to the PASC increased up to 60%, sending bands 2, 3 and 8 into saturation (band 1 was saturated at launch). The short-wave infrared (SWIR) band individual detectors have shown variations up to - 20% in response to the PASC. The variation is different for each detector. After the first six months, the responses to the PASC have become more stable, with much of the variation related to the within-scan position of the solar image. Overall results to date for all calibrators and comparisons with vicarious calibrations indicate that most of the response variations have been due to the calibrators themselves and suggest that the instrument has been stable with changes in response of less than 0.5% y m 1 .

International Journal of Remote Sensing↗

Landsat-5 TM reflective-band absolute radiometric calibration

The Landsat-5 Thematic Mapper (TM) sensor provides the longest running continuous dataset of moderate spatial resolution remote sensing imagery, dating back to its launch in March 1984. Historically, the radiometric calibration procedure for this imagery used the instrument's response to the Internal Calibrator (IC) on a scene-by-scene basis to determine the gain and offset of each detector. Due to observed degradations in the IC, a new procedure was implemented for U.S.-processed data in May 2003. This new calibration procedure is based on a lifetime radiometric calibration model for the instrument's reflective bands (1-5 and 7) and is derived, in part, from the IC response without the related degradation effects and is tied to the cross calibration with the Landsat-7 Enhanced Thematic Mapper Plus. Reflective-band absolute radiometric accuracy of the instrument tends to be on the order of 7% to 10%, based on a variety of calibration methods.

IEEE Transactions on Geoscience and Remote Sensing↗

Landsat-7 ETM+ on-orbit reflective-band radiometric stability and absolute calibration

Launched in April 1999, the Landsat-7 Enhanced Thematic Mapper Plus (ETM+) instrument is in its sixth year of operation. The ETM+ instrument has been the most stable of any of the Landsat instruments. To date, the best onboard calibration source for the reflective bands has been the Full Aperture Solar Calibrator, a solar-diffuser-based system, which has indicated changes of between 1% to 2% per year in the ETM+ gain for bands 1-4 and 8 and less than 0.5%/year for bands 5 and 7. However, most of this change is believed to be caused by changes in the solar diffuser panel, as opposed to a change in the instrument's gain. This belief is based partially on vicarious calibrations and observations of "invariant sites", hyperarid sites of the Sahara and Arabia. Weighted average slopes determined from these datasets suggest changes of 0.0% to 0.4% per year for bands 1-4 and 8 and 0.4% to 0.5% per year for bands 5 and 7. Absolute calibration of the reflective bands of the ETM+ is consistent with vicarious observations and other sensors generally at the 5% level, though there appear to be some systematic differences.

IEEE Transactions on Geoscience and Remote Sensing↗

Landsat-7 ETM+ on-orbit reflective-band radiometric characterization

The Landsat-7 Enhanced Thematic Mapper Plus (ETM+) has been and continues to be radiometrically characterized using the Image Assessment System (IAS), a component of the Landsat-7 Ground System. Key radiometric properties analyzed include: overall, coherent, and impulse noise; bias stability; relative gain stability; and other artifacts. The overall instrument noise is characterized across the dynamic range of the instrument during solar diffuser deployments. Less than 1% per year increases are observed in signal-independent (dark) noise levels, while signal-dependent noise is stable with time. Several coherent noise sources exist in ETM+ data with scene-averaged magnitudes of up to 0.4 DN, and a noise component at 20 kHz whose magnitude varies across the scan and peaks at the image edges. Bit-flip noise does not exist on the ETM+. However, impulse noise due to charged particle hits on the detector array has been discovered. The instrument bias is measured every scan line using a shutter. Most bands show less than 0.1 DN variations in bias across the instrument lifetime. The panchromatic band is the exception, where the variation approaches 2 DN and is related primarily to temperature. The relative gains of the detectors, i.e., each detector's gain relative to the band average gain, have been stable to /spl plusmn/0.1% over the mission life. Two exceptions to this stability include band 2 detector 2, which dropped about 1% in gain about 3.5 years after launch and stabilized, and band 7 detector 5, which has changed several tenths of a percent several times since launch. Memory effect and scan-correlated shift, a hysteresis and a random change in bias between multiple states, respectively, both of which have been observed in previous Thematic Mapper sensors, have not been convincingly found in ETM+ data. Two artifacts, detector ringing and "oversaturation", affect a small amount of ETM+ data.

IEEE Transactions on Geoscience and Remote Sensing↗

Landsat-7 ETM+ radiometric stability and absolute calibration

Launched in April 1999, the Landsat-7 ETM+ instrument is in its fourth year of operation. The quality of the acquired calibrated imagery continues to be high, especially with respect to its three most important radiometric performance parameters: reflective band instrument stability to better than ??1%, reflective band absolute calibration to better than ??5%, and thermal band absolute calibration to better than ??0.6 K. The ETM+ instrument has been the most stable of any of the Landsat instruments, in both the reflective and thermal channels. To date, the best on-board calibration source for the reflective bands has been the Full Aperture Solar Calibrator, which has indicated changes of at most -1.8% to -2.0% (95% C.I.) change per year in the ETM+ gain (band 4). However, this change is believed to be caused by changes in the solar diffuser panel, as opposed to a change in the instrument's gain. This belief is based partially on ground observations, which bound the changes in gain in band 4 at -0.7% to +1.5%. Also, ETM+ stability is indicated by the monitoring of desert targets. These image-based results for four Saharan and Arabian sites, for a collection of 35 scenes over the three years since launch, bound the gain change at -0.7% to +0.5% in band 4. Thermal calibration from ground observations revealed an offset error of +0.31 W/m 2 sr um soon after launch. This offset was corrected within the U. S. ground processing system at EROS Data Center on 21-Dec-00, and since then, the band 6 on-board calibration has indicated changes of at most +0.02% to +0.04% (95% C.I.) per year. The latest ground observations have detected no remaining offset error with an RMS error of ??0.6 K. The stability and absolute calibration of the Landsat-7 ETM+ sensor make it an ideal candidate to be used as a reference source for radiometric cross-calibrating to other land remote sensing satellite systems.

Conference Paper↗

Landsat-7 ETM+ radiometric calibration: Two years on-orbit

Landsat-7 has been in orbit for 2 years as of April 15, 2001 and operationally providing calibrated data products for 2 years as of June 28, 2001. A radiometric calibration team consisting of scientists and analysts from the Landsat Project Science Office, the Landsat-7 Image Assessment System and four universities evaluates the calibration based on on-board and ground-look (vicarious) calibration methodologies. The results are assembled and compared semi-annually and the calibration parameter files are adjusted as necessary. To date the combined results for the reflective bands have not shown any change from pre-launch values. The pre-launch values continue to be used for data processing, with the uncertainty estimated at less than 5%. In the thermal band, the vicarious calibration results indicated a 0.31 W/m/sup 2/ sr /spl mu/m bias in the calibration. This bias results in the ETM+ derived temperatures being about 3K high. The calibration parameter file was updated October 1, 2000 to remove this bias, however the U.S. Landsat Product Generation System (LPGS) software required modification that was not incorporated until December 20, 2000. All LPGS data products generated since this date have the correct thermal band calibration, regardless of image acquisition date, with uncertainties at approximately the 1% level.

Conference Paper↗